Abstract
Little is known about the accuracy of intraesophageal pH monitoring, ie, do false negatives occur? This study was designed based on the assumption that if two pH probes are placed equidistant from a distal acid source, and the probes measure different values, then one probe might miss acid if it were there alone. We performed 24-hr intraesophageal pH monitoring in 10 patients with gastroesophageal reflux disease, simultaneously placing two pediatric pH probes (probes A and B) 5 cm proximal to the LES in such close proximity that one would expect only small differences in the acid exposure measured by the probes. However, the amount of acid exposure measured by probes A and B differed greatly in some instances. Individual probe performance does not alone account for these differences, since, when pH recordings were reviewed, one probe was as likely to miss acid exposure as the other. These differences would result in a change in the clinical diagnosis in two of 10 patients. Variability in the number of episodes was great and occurred even when there was concordance in percent acid exposure. These data raise questions about the absolute accuracy of the pH probe as the gold standard in the measurement of gastroesophageal reflux.
Similar content being viewed by others
References
Richter JE, Castell DO: Gastroesophageal reflux. Pathogenesis, diagnosis, and therapy. Ann Intern Med 97:93–103, 1982
Johnson LF: New concepts and methods in the study and treatment of gastroesophageal reflux disease. Med Clin North Am 65:1195–1220, 1981
Johnson LF: 24-hour pH monitoring in the study of gastroesophageal reflux. J Clin Gastroenterol 2:387–399, 1980
Ward BW, Wu WC, Richter JE, Lui KW, Castell DO: Ambulatory 24-hour esophageal pH monitoring. J Clin Gastroenterol 8(suppl 1):59–67, 1986
McLauchlan G, Rawlings JM, Lucas ML: Electrodes for 24 hour pH monitoring-a comparative study. Gut 28:935–939, 1987
Rovelstad RA, Owen CA, Magath TB: Factors influencing the continuous recording ofin situ pH of gastric and duodenal contents. Gastroenterology 20:609–624, 1952
Bircher J, Mann CV, Carlson, Code CF, Rovelstad RA: Intraluminal and juxtamucosal duodenal pH. Gastroenterology 48:472–477, 1965
Quigley EMM, Turnberg LA: pH of the microclimate lining human gastric and duodenal mucosain vivo. Gastroenterology 92:1876–1884, 1987
Orr WC, Johnson LF, Robinson MG: Effect of sleep on esophageal peristalsis and acid clearance. Gastroenterology 86:814–819, 1984
Helm JF, Dodds WJ, Pelc LR: Effect of esophageal emptying and saliva on clearance of acid from the esophagus. N Engl J Med 310:284–288, 1984
Schlesinger PK, Donahue PE, Schmid B, Layden TJ: Limitations of 24-hour intraesophageal pH monitoring in the hospital setting. Gastroenterology 89:797–804, 1985
Herrera JL, Simpson JK, Maydonovitch CL, Wong RK: Comparison of stationary vs ambulatory 24-hour pH monitoring recording systems. Dig Dis Sci 33:385–388, 1988
Author information
Authors and Affiliations
Additional information
Dr. Yuan performed this work as senior student research project.
Rights and permissions
About this article
Cite this article
Murphy, D.W., Yuan, Y. & Castell, D.O. Does the intraesophageal pH probe accurately detect acid reflux?. Digest Dis Sci 34, 649–656 (1989). https://doi.org/10.1007/BF01540333
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01540333