Outcomes of Unilateral Versus Bilateral Use of the Iliac Branch Endoprosthesis for Elective Endovascular Treatment of Aorto-iliac Aneurysms

  • Clinical Investigation
  • Arterial Interventions
  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose

The aim of the study was to evaluate outcomes after bilateral implantation of the Gore Excluder Iliac Branch Endoprosthesis (IBE) versus those achieved after unilateral implantation.

Methods

All consecutive patients electively treated in a single center for aorto-iliac aneurysm using the IBE device between January 1, 2014, and December 31, 2018, were reviewed. Early outcome measures were technical success, 30 days or in-hospital mortality, and major adverse events (MAE). Late outcome measures were survival, internal iliac artery (IIA) patency, and freedom from IIA branch instability.

Results

A total of 74 patients (97% males, mean age 74 ± 7 years) were included. Thirteen patients (17%) received bilateral IBE implantation for a total of 85 vessels evaluated. The technical success rate was 97% and was not significantly different between the two groups (p = .32). Two patients died within 30 days, both in the unilateral group (p = 1). No significant differences were seen in the rates of 30 days MAE (p = .10). At one year, the overall survival rate was 95 ± 2% vs 90 ± 3% in the unilateral and bilateral group, respectively (Log-rank = .05). There were no differences in 1-year primary and secondary patency rates between groups (Log-rank = .75 and Log-rank = .34, respectively). Freedom from IIA branch instability at one year was also not significantly different (unilateral: 94 ± 3% vs. bilateral: 82 ± 9%, Log-rank = .22)..

Conclusions

Bilateral IBE use for elective endovascular treatment of aorto-iliac aneurysms appears safe and feasible and may achieve satisfactory short-term and mid-term outcomes. Bilateral IBE use should be employed judiciously in the context of a comprehensive risk/benefit evaluation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Li Y, Hu Z, Zhang J, Zheng H, Wang J, Chen Z, et al. Iliac aneurysms treated with endovascular iliac branch device: a systematic review and meta-analysis. Ann Vasc Surg. 2019;56:303–16.

    Article  Google Scholar 

  2. D’Oria M, Mendes B, Bews K, Hanson K, Johnstone J, Shuja F, et al. Perioperative outcomes after use of iliac branch devices compared with hypogastric occlusion or open surgery for elective treatment of aortoiliac aneurysms in the NSQIP database. Ann Vasc Surg. 2020;62:35–44.

    Article  Google Scholar 

  3. Schneider DB, Matsumura JS, Lee JT, Peterson BG, Chaer RA, Oderich GS. Prospective, multicenter study of endovascular repair of aortoiliac and iliac aneurysms using the Gore Iliac Branch Endoprosthesis. J Vasc Surg. 2017;66:775–85.

    Article  Google Scholar 

  4. D’Oria M, Pitoulias GA, Torsello GF, Pitoulias AG, Fazzini S, Masciello F, et al. Bilateral use of iliac branch devices for aortoiliac aneurysms is safe and feasible, and procedural volume does not seem to affect technical or clinical effectiveness: early and midterm results from the pELVIS international multicentric registry. J Endovasc Ther. 2021;28:585–92.

    Article  Google Scholar 

  5. D’Oria M, Tenorio ER, Oderich GS, DeMartino RR, Kalra M, Shuja F, et al. Outcomes after standalone use of Gore Exclude Iliac Branch Endoprosthesis for endovascular repair of isolated iliac artery aneurysms. Ann Vasc Surg. 2020;67:158–70.

    Article  Google Scholar 

  6. D’Oria M, Oderich GS, Tenorio ER, Karkkainen JM, Mendes BC, DeMartino R. Safety and efficacy of totally percutaneous femoral access for fenestrated-branched endovascular aortic repair of pararenal-thoracoabdominal aortic aneurysms. Cardiovasc Intervent Radiol. 2020;43:547–55.

    Article  Google Scholar 

  7. Dawson DL, Sandri GA, Tenorio ER, Oderich GS. Up-and-over technique for implantation of iliac branch devices after prior aortic endograft repair. J Endovasc Ther. 2018;25:21–7.

    Article  Google Scholar 

  8. Tenorio ER, Oderich GS, Sandri GA, Karkkainen JM, Kalra M, DeMartino RR, et al. Outcomes of an iliac branch endoprosthesis using an “up and over” technique for endovascular repair of failed bifurcated grafts. J Vasc Surg. 2019;70:497–508.

    Article  Google Scholar 

  9. Chaikof EL, Blankensteijn JD, Harris PL, White GH, Zarins CK, Bernhard VM, et al. Reporting standards for endovascular aortic aneurysm repair. J Vasc Surg. 2002;35:1048–60.

    Article  Google Scholar 

  10. Lepidi S, D’Oria M. Long term benefits of EVAR in the modern era: the importance of looking at stent graft durability without forgetting the right pathophysiology. Eur J Vasc Endovasc Surg. 2020;60:956–7.

    Article  Google Scholar 

  11. Chaikof E, Dalman RL, Eskandari M, Jackson BM, Lee WA, Mansour MA, et al. The Society for Vascular Surgery practice guidelines on the care of patients with an abdominal aortic aneurysm. J Vasc Surg. 2018;67:2–77.

    PubMed  Google Scholar 

  12. Wanhainen A, Verzini F, Van Herzeele I, Allaire E, Bown M, Cohnert T, et al. European Society for Vascular Surgery (ESVS) 2019 clinical practice guidelines on the management of abdominal aorto-iliac artery aneurysms. Eur J Vasc Endovasc Surg. 2019;57:8–93.

    Article  Google Scholar 

  13. Mendes BC, Oderich GS, Sandri GA, Johnstone JK, Shuja F, Kalra M, et al. Comparison of perioperative outcomes of patients with iliac aneurysms treated by open surgery or endovascular repair with iliac branch endoprosthesis. Ann Vasc Surg. 2019;60:76–84.

    Article  Google Scholar 

  14. Kliewer M, Plimon M, Taher F, Walter C, Hirsch K, Falkensammer J, et al. Endovascular treatment of hypogastric artery aneurysms. J Vasc Surg. 2019;70:1107–14.

    Article  Google Scholar 

  15. Pini R, Faggioli G, Indelicato G, Gallitto E, Mascoli C, Stella A, et al. Early and late outcome of common iliac aneurysms treated by flared limbs or iliac branch devices during endovascular aortic repair. J Vasc Intervent Radiol. 2019;30:503–10.

    Article  Google Scholar 

  16. Giosdekos A, Antonopoulos CN, Sfyroeras G, Moulakakis KG, Tsilimparis N, Kakisis JD, et al. The use of iliac branch devices for preservation of flow in internal iliac artery during endovascular aortic aneurysm repair. J Vasc Surg. 2020;71:2133–44.

    Article  Google Scholar 

  17. de Marino PM, Botos B, Kouvelos G, Verhoeven ELG, Katsargyris A. Use of bilateral cook Zenith Iliac Branch Devices to preserve internal iliac artery flow during endovascular aneurysm repair. Eur J Vasc Endovasc Surg. 2019;57:213–9.

    Article  Google Scholar 

  18. Maldonado TS, Mosquera NJ, Lin P, Bellosta R, Barfield M, Mouossa A, et al. Gore Iliac Branch Endoprosthesis for treatment of bilateral common iliac artery aneurysms. J Vasc Surg. 2018;68:100–8.

    Article  Google Scholar 

  19. D’Oria M, Mastrorilli D, DeMartino R, Lepidi S. Current status of endovascular preservation of the internal iliac artery with iliac branch devices. Cardiovasc Intervent Radiol. 2019;42:935–48.

    Article  Google Scholar 

  20. Schneider DB, Milner R, Heyligers JMM, Chakfe N, Matsumura J. Outcomes of the GORE Iliac Branch Endoprosthesis in clinical trial and real-world registry settings. J Vasc Surg. 2019;69:367–77.

    Article  Google Scholar 

  21. Eagleton MJ, Shah S, Petkosevek D, Mastracci TM, Greenberg RK. Hypogastric and subclavian artery patency affects onset and recovery of spinal cord ischemia associated with aortic endografting. J Vasc Surg. 2014;59:89–95.

    Article  Google Scholar 

  22. Simonte G, Parlani G, Farchioni L, Isernia G, Cieri E, Lenti M, et al. Lesson learned with the use of iliac branch devices: single centre 10 year experience in 157 consecutive patients. Eur J Vasc Endovasc Surg. 2017;54:95–103.

    Article  CAS  Google Scholar 

  23. Delay C, Deglise S, Lejay A, Georg Y, Roussin M, Schaeffer M, et al. Zenith bifurcated iliac side branch device: midterm results and assessment of risk factors for intraoperative thrombosis. Ann Vasc Surg. 2017;41:141–50.

    Article  Google Scholar 

  24. Tenorio ER, Oderich GS, Sandri GA, Ozbek P, Karkkainen JM, Macedo TA, et al. Impact of onlay fusion and cone beam computed tomography on radiation exposure and technical assessment of fenestrated-branched endovascular aortic repair. J Vasc Surg. 2019;69:1045–58.

    Article  Google Scholar 

  25. Bosanquet DC, Wilcox C, Whitehurst L, Cox A, Williams IM, Twine CP, British Society of Endovascular Therapy (BSET). Systematic review and meta-analysis of the effect of internal iliac artery occlusion for patients undergoing EVAR. Eur J Vasc Endovasc Surg. 2017;53:534–48.

    Article  CAS  Google Scholar 

  26. Kouvelos GN, Katsargyris A, Antoniou GA, Oikonomou K, Verhoeven EL. Outcome after interruption or preservation of internal iliac artery flow during endovascular repair of abdominal aorto-iliac aneurysms. Eur J Vasc Endovasc Surg. 2016;52:621–34.

    Article  CAS  Google Scholar 

  27. Bosiers MJ, Panuccio G, Bisdas T, Stachmann A, Donas KP, Torsello G, et al. Longer bridging stent grafts in iliac branch endografting doesn’t worsen outcome and expand its applicability, even in concomitant diseased hypogastric arteries. J Cardiovasc Surg. 2020;61:191–5.

    Google Scholar 

  28. D’Oria M, Pipitone M, Sgorlon G, Chiarandini S, Rotelli A, Griselli F. Endovascular exclusion of hypogastric aneurysms using distal branches of the internal iliac artery as landing zone: a case series. Ann Vasc Surg. 2018;46:369.e13-369.e18.

    Article  Google Scholar 

  29. D’Oria M, Tenorio ER, Oderich GS, Mendes BC, Kalra M, Shuja F, et al. Outcomes of the Gore Excluder Iliac Branch Endoprosthesis using divisional branches of the internal iliac artery as distal landing zone. J Endovasc Ther. 2020;27:316–27.

    Article  Google Scholar 

  30. Jerkku T, Mohammed WM, Kapetanios D, Czihal M, Tsilimparis N, Banafsche R. Extension of iliac branch device repair into the superior gluteal artery is a safe and effective maneuver. Ann Vasc Surg. 2020;62:195–205.

    Article  Google Scholar 

  31. Dueppers P, Duran M, Floros N, Schelzig G, Wagenhauser MU, Oberhuber A. The JOTEC iliac branch device for exclusion of hypogastric artery aneurysms: ABRAHAM study. J Vasc Surg. 2019;70:748–55.

    Article  Google Scholar 

  32. Karthikesalingam A, Hinchliffe RJ, Malkawi AH, Holt PJ, Loftus IM, Thompson MM. Morphological suitability of patients with aortoiliac aneurysms for endovascular preservation of the internal iliac artery using commercially available iliac branch graft devices. J Endovasc Ther. 2010;17:163–71.

    Article  Google Scholar 

  33. Gray D, Shahverdyan R, Jakobs C, Brunkwall J, Gawenda M. Endovascular aneurysm repair of aortoiliac aneurysms with an iliac side-branched stent-graft: studying the morphological applicability of the Cook device. Eur J Vasc Endovasc Surg. 2015;49:283–8.

    Article  CAS  Google Scholar 

  34. Pearce BJ, Varu VN, Glocker R, Novak Z, Jordan WD, Lee JT. Anatomic suitability of aortoiliac aneurysms for next generation branched systems. Ann Vasc Surg. 2015;29:69–75.

    Article  Google Scholar 

  35. e Melo RG, Fenelli C, Prendes CF, Oz T, Stavroulakis K, Rantner B, et al. A cross-sectional study on the anatomical feasibility of iliac side branch grafts in a real-world setting. J Vasc Surg. 2022. https://doi.org/10.1016/j.jvs.2022.02.056.

    Article  Google Scholar 

Download references

Funding

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Mario D’Oria.

Ethics declarations

Conflict of interest

G.S.O. has received consulting fees and grants from W.L. Gore (all paid to Mayo Clinic with no personal income).

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

D’Oria, M., Tenorio, E.R., Oderich, G.S. et al. Outcomes of Unilateral Versus Bilateral Use of the Iliac Branch Endoprosthesis for Elective Endovascular Treatment of Aorto-iliac Aneurysms. Cardiovasc Intervent Radiol 45, 939–949 (2022). https://doi.org/10.1007/s00270-022-03166-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00270-022-03166-3

Keywords

Navigation