Abstract
Epidrum® is an optimal pressure, loss of resistance device for identifying the epidural space. We investigated the usefulness of Epidrum versus the loss of resistance or hanging drop techniques while performing epidural anesthesia. Eighty adult patients who were scheduled for elective surgery under lumbar epidural anesthesia were randomized into two groups. The first group (Epidrum group) consisted of 40 adult patients who were scheduled for epidural anesthesia using Epidrum. The second group (control group) consisted of 40 adult patients who were scheduled for epidural anesthesia using the loss of resistance or hanging drop technique. We recorded the time required to identify the epidural space and outcomes of epidural catheterization. The attending anesthesiologists were also questioned regarding the ease of control of the Tuohy needle and of epidural space identification with each method. The time required to perform epidural anesthesia was significantly shorter in the Epidrum group than in the control group [28 s (10–76) vs. 90 s (34–185); median (interquartile range)] (p < 0.05). Tuohy needle control was significantly easier in the Epidrum group than in the control group (p < 0.05). Epidrum is useful for performing epidural anesthesia quickly while obtaining good Tuohy needle control.
References
Brown DL. Spinal, epidural, and caudal anesthesia. In: Miller RD, editor. Miller’s anesthesia. 7th ed. Philadelphia: Churchill Livingstone Elsevier; 2010. p. 1611–38.
Lechner TJM, van Wijk MGF, Maas AJJ. Clinical results with a new acoustic device to identify the epidural space. Anaesthesia. 2002;57:768–72.
Zarzur E. Genesis of the ‘true’ negative pressure in the lumbar epidural space: a new hypothesis. Anaesthesia. 1984;39:1101–4.
Liu S, Carpenter RL, Neal JM. Epidural anesthesia and analgesia: their role in postoperative outcome. Anesthesiology. 1995;82:1474–506.
Eappen S, Blinn A, Segal S. Incidence of epidural catheter replacement in parturients: a retrospective chart review. Int J Obstet Anesth. 1998;7:220–5.
Jacob S, Tierney E. A dual technique for identification of the epidural space. Anaesthesia. 1997;52:141–3.
Lechner TJM, van Wijk MGF, Maas AJJ. Clinical results with the acoustic puncture assist device, a new acoustic device to identify the epidural space. Anesth Analg. 2003;96:1183–7.
Conflict of interest
This research was supported solely by institutional and/or departmental sources. None of the authors have any financial interests in products related to this study.
Author information
Authors and Affiliations
Corresponding author
Additional information
This paper was presented in part at the 29th annual meeting of the Japan Society of Clinical Anesthesia, Hamamatsu, Japan, 29–31 October 2009 and at the annual meeting of the American Society of Anesthesiologists, San Diego, CA, USA, 16–20 October 2010.
About this article
Cite this article
Sawada, A., Kii, N., Yoshikawa, Y. et al. Epidrum®: a new device to identify the epidural space with an epidural Tuohy needle. J Anesth 26, 292–295 (2012). https://doi.org/10.1007/s00540-011-1278-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-011-1278-1