Three-dimensional microanatomy of human nipple visualized by X-ray dark-field computed tomography

  • Preclinical study
  • Published:
Breast Cancer Research and Treatment Aims and scope Submit manuscript

Abstract

Purpose

The three-dimensional (3D) structure of the human nipple has not been fully clarified. However, its importance has increased in recent years because it has become common practice to preoperatively explore the spread of breast cancer to the nipple with needle biopsy, ductoscopy, and/or ductal lavage for nipple-sparing mastectomy. Here, we demonstrated that X-ray dark-field computed tomography (XDFI-CT) is a powerful tool for reconstructing the 3D distribution pattern of human lactiferous ducts non-destructively, without contrast agent, and with high tissue contrast.

Methods

Nipples amputated from mastectomy specimens of 51 patients with breast cancer were visualized three-dimensionally by XDFI-CT. First, CT images and conventionally stained tissue sections were compared to demonstrate that XDFI-CT provides 3D anatomical information. Next, the number of ducts in the nipple and the number of ducts sharing an ostium near the tip of the nipple were measured from the volume set of XDFI-CT. Finally, the 3D distribution pattern of the ducts was determined.

Results

XDFI-CT can provide images almost equivalent to those of low-magnification light microscopy of conventional hematoxylin–eosin-stained histological sections. The mean number of ducts in all cases was 28.0. The total number of ducts sharing an ostium near the tip of the nipple was 525 of 1428. The 3D distribution patterns of the ducts were classified into three types that we defined as convergent (22%), straight (39%), or divergent (39%).

Conclusions

XDFI-CT is useful for exploring the microanatomy of the human nipple and might be used for non-invasive nipple diagnosis in the future.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

Data availability

The datasets generated and analyzed during the current study are available from the corresponding author on reasonable request.

References

  1. Going JJ, Moffat DF (2004) Escaping from Flatland: clinical and biological aspects of human mammary duct anatomy in three dimensions. J Pathol 203(1):538–544

    Article  Google Scholar 

  2. Rusby JE, Brachtel EF, Michaelson JS et al (2007) Breast duct anatomy in the human nipple: three-dimensional patterns and clinical implications. Breast Cancer Res Treat 106(2):171–179

    Article  Google Scholar 

  3. Taneri F, Kurukahvecioglu O, Akyurek N et al (2006) Microanatomy of milk ducts in the nipple. Eur Surg Res 38(6):545–549

    Article  Google Scholar 

  4. Rusby JE, Smith BL, Gui GP (2010) Nipple-sparing mastectomy. Br J Surg 97(3):305–316

    Article  CAS  Google Scholar 

  5. Chang RYK, Cheung PSY (2017) Nipple preservation in breast cancer associated with nipple discharge. Would J Surg 41(1):176–183

    Article  Google Scholar 

  6. Makita M, Akiyama F, Gomi N et al (2016) Mammary ductoscopy and watchful follow-up substitute microdochectomy in patients with bloody nipple discharge. Breast Cancer 23(2):242–251

    Article  Google Scholar 

  7. Yaegashi H, Takahashi T, Kawasaki M (1987) Microcomputer-aided reconstruction: a system designed for the study of 3-D microstructure in histology and histopathology. J Microsc 146(1):55–65

    Article  CAS  Google Scholar 

  8. Ando M, Maksimenko A, Sugiyama H et al (2002) Simple X-ray dark-and bright-field imaging using achromatic Laue optics. Jpn J Appl Phys 41(9A):L1016–L1018

    Article  CAS  Google Scholar 

  9. Sunaguchi N, Yuasa T, Huo Q et al (2010) X-ray refraction-contrast computed tomography images using dark-field imaging optics. Appl Phys Lett 97(15):153701–153703

    Article  Google Scholar 

  10. Snigirev A, Snigireva I, Kohn V, Kuznetsov S et al (1995) On the possibilities of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation. Rev Sci Instrum 66(12):5486–5492

    Article  CAS  Google Scholar 

  11. Momose A, Takeda T, Itai Y, Yoneyama A et al (1998) Phase-contrast tomographic imaging using an X-ray interferometer. J Synchrotron Radiat 5(3):309–314

    Article  CAS  Google Scholar 

  12. Chapman D, Thomlinson W, Johnston RE et al (1997) Diffraction enhanced x-ray imaging. Phys Med Biol 42(11):2015–2025

    Article  CAS  Google Scholar 

  13. Momose A, Kawamoto S, Koyama I et al (2003) Demonstration of X-ray Talbot interferometry. Jpn J Appl Phys 42(7B):L866–L868

    Article  CAS  Google Scholar 

  14. Ando M et al (2016) Dark-field imaging: recent developments and potential clinical applications. Phys Med 32:1801–1812

    Article  Google Scholar 

  15. Sunaguchi N, Shimao D, Ichihara S et al (2019) Three-dimensional reconstruction of human nipple using refraction-contrast x-ray computed tomography. AIP Conf Proc 2054:050010

    Article  Google Scholar 

  16. Sunaguchi N, Yuasa T, Gupta R et al (2015) An efficient reconstruction algorithm for differential phase-contrast tomographic images from a limited number of views. Appl Phys Lett 107(25):253701–253704

    Article  Google Scholar 

  17. Grubbs FE (1950) Sample criteria for testing outlying observations. Ann Math Stat 21(1):27–58

    Article  Google Scholar 

  18. Young WA, Degnim AC, Hoskin TL et al (2019) Outcomes of%3e 1300 nipple-sparing mastectomies with immediate reconstruction: the impact of expanding indications on complications. Ann Surg Oncol 26(10):3115–3123

    Article  Google Scholar 

  19. Lee CH, Cheng MH, Wu CW et al (2019) Nipple-sparing mastectomy and immediate breast reconstruction after recurrence from previous breast conservation therapy. Ann Plast Surg 82(1S):S95–102

    Article  CAS  Google Scholar 

  20. Mallon P, Feron JG, Couturaud B (2013) The role of nipple-sparing mastectomy in breast cancer: a comprehensive review of the literature. Plast Reconstr Surg 131(5):969–984

    Article  CAS  Google Scholar 

  21. Crowe JP Jr, Kim JA, Yetman R et al (2004) Nipple-sparing mastectomy: technique and results of 54 procedures. Arch Surg 139(2):148–150

    Article  Google Scholar 

  22. Antoniassi M, Conceição ALC, Poletti ME (2012) Study of electron densities of normal and neoplastic human breast tissues by Compton scattering using synchrotron radiation. Appl Radiat Isot 70(7):1351–1354

    Article  CAS  Google Scholar 

  23. Ichihara S, Ando M, Maksimenko A et al (2008) 3-D reconstruction and virtual ductoscopy of high-grade ductal carcinoma in situ of the breast with casting type calcifications using refraction-based X-ray CT. Virchows Arch 452(1):41–47

    Article  Google Scholar 

  24. Tan MP, Tot T (2018) The sick lobe hypothesis, field cancerisation and the new era of precision breast surgery. Gland Surg 7(6):611–618

    Article  Google Scholar 

  25. Going JJ, Mohun TJ (2006) Human breast duct anatomy, the‘sick lobe’ hypothesis and intraductal approaches to breast cancer. Breast Cancer Res Treat 97:285–291

    Article  Google Scholar 

  26. Bech M, Bunk O, David C et al (2009) Hard X-ray phase-contrast imaging with the compact light source based on inverse Compton X-rays. J Synchrotron Radiat 16(1):43–47

    Article  Google Scholar 

  27. Kuroda R, Toyokawa H, Yasumoto M et al (2011) Quasi-monochromatic hard X-ray source via laser Compton scattering and its application. Nucl Instrum Methods Phys Res 637(1):S183–S186

    Article  CAS  Google Scholar 

  28. Han Y, Ye JC (2018) Framing U-Net via deep convolutional framelets: application to sparse-view CT. IEEE Trans Med Imaging 37(6):1418–1429

    Article  Google Scholar 

  29. Huang F, Hartwich TM, Rivera-Molina FE et al (2013) Video-rate nanoscopy using sCMOS camera–specific single-molecule localization algorithms. Nat Methods 10(7):653

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The Si crystal with the imaging optics was manufactured at the Photon Factory Crystal Optics Processing Laboratory within KEK, Tsukuba, Japan by Dr. H. Sugiyama and Mr. T. Sasaya. The X-ray cameras for the XDFI-CT system were borrowed from Dr. K. Hirano and Dr. K. Hyodo. The experiments were performed as part of four research projects (2008S2002, 2016G0625, 2012G562, and 2015G597) funded by the KEK. This research was partially supported by a Grant-in-Aid for Scientific Research (grant numbers 16K01369, 16K08654, 22591353, 15H01129, 26286079, and 18K13765) from the Japanese Ministry of Education, Culture, Sports, Science and Technology.

Author information

Authors and Affiliations

Authors

Contributions

NS, DS, RO, AW, and KN obtained the duct positions from the CT data. NS performed the subjective classification of the duct structure. NS developed the mathematical method to classify the types of 3D duct structures. SI and RN prepared all nipple samples used in this research. NS reconstructed the CT images of all samples. SI prepared the histological sections of each nipple sample. NS wrote the main text of the manuscript, and SI, SD, RN, TY, and MA reviewed the manuscript.

Corresponding author

Correspondence to Naoki Sunaguchi.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures used in this research were approved by the Ethical Committee of Nagoya University and Nagoya Medical Center.

Informed consent

With regard to the appropriate procedure for collecting data from existing tissue samples, we adopted an opt-out system instead of informed consent. Our website and bulletin boards announced that we were collecting data from existing tissue samples for scientific research in the present study. The patients’ personal information is protected, but the data will be used automatically unless the patients actively dissent.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Sunaguchi, N., Shimao, D., Yuasa, T. et al. Three-dimensional microanatomy of human nipple visualized by X-ray dark-field computed tomography. Breast Cancer Res Treat 180, 397–405 (2020). https://doi.org/10.1007/s10549-020-05574-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10549-020-05574-w

Keywords

Navigation