Abstract
A new relative-humidity setup was developed for calibrating sensors in the temperature range from \(-40\,^{\circ }\mathrm{C}\) up to \(180\,^{\circ }\mathrm{C}\) and at pressures down to 700 hPa and up to 0.5 MPa. The setup is based on the chamber-in-chamber model: a small additional chamber is positioned inside a climatic chamber. While the climatic chamber is used to generate the air temperature, a pre-conditioned gas from outside the climatic chamber delivers the required humidity in the new pressure chamber. Validation of the setup at atmospheric pressure showed relative-humidity uncertainties of 0.2 %rh at 5 %rh over the whole temperature range and 0.4 %rh at 95 %rh for temperatures above \(0\,^{\circ }\mathrm{C}\). Below \(0\,^{\circ }\mathrm{C}\), the maximum uncertainty increases to 0.9 %rh due to the influence of the temperature homogeneity. The temperature uncertainty of the new setup is between \(0.10\,^{\circ }\mathrm{C}\) and \(0.21\,^{\circ }\mathrm{C}\). Five commercially available relative-humidity sensors, of different type and manufacturer and all suitable for high temperatures, were calibrated in the new setup. The measurements showed deviations outside the stated specifications of the manufacturer and the need of traceable calibration facilities.
Similar content being viewed by others
References
R. Mokhutshoane, J. Nielsen, C. Barendregt, M.J. de Groot, “The Set-Up and Characterization of a Calibration System for Dewpoint Meters and Relative Humidity Sensors,” in Proceedings of 4th ISHM 2002 (HSU, ITRI, Taipei, 2002), pp. 493–499
D. Sonntag, Z. Meteorol. 40, 340 (1990)
W. Wagner, A. Pruss, J. Phys. Chem. Ref. Data 22, 783 (1993)
R. Hardy, “ITS-90 Formulations for Vapor Pressure, Frostpoint Temperature, Dewpoint Temperature and Enhancement Factors in the Range \(-100\) to +100C,” in Papers and Abstracts from the Third International Symposium on Humidity and Moisture (NPL, London, 1998), pp. 214–222
N. Böse, H. Mitter, “Calculation and Measurement of Relative Humidity at Temperatures Greater than \(100^{\circ }{\rm C}\),” in Proceedings of TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by D. Zvizdić, L.G. Bermanec, T. Veliki, T. Stašić (FSB/LPM, Zagreb, 2004), pp. 689–695
J. Nielsen, C. Barendregt, “The Use of Thermistors for Establishing the Temperature Conditions in a Climatic Chamber,” in Proceedings of TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by D. Zvizdić, L.G. Bermanec, T. Veliki, T. Stašić (FSB/LPM, Zagreb, 2004), pp. 781–786
J. van Geel, R. Bosma, J. van Wensveen, A. Peruzzi, “Thermistors Used in Climatic Chamber at High Temperature and Humidity,” Int. J. Thermophys. (submitted)
“Calibration of Climatic Chambers, Guideline DKD-R 5-7,” edn. 07/2004, English translation 02/2009
Acknowledgments
This work was funded by the Ministry of Economic Affairs, The Netherlands.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Bosma, R., Peruzzi, A. Climatic Chamber for Temperatures up to \(180\,^{\circ }\mathrm{C}\) and Pressures up to 0.5 MPa. Int J Thermophys 35, 738–747 (2014). https://doi.org/10.1007/s10765-014-1596-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-014-1596-2