Abstract
This paper compares the healing of supported knitted Dacron prostheses implanted in the descending thoracic aorta and in the subcutaneous carotid-femoral positions in each of 10 dogs. The descending thoracic aorta prostheses were 6 cm long and the carotid-femoral prostheses averaged 76 cm in length. The healing of four porous clinical axillofemoral grafts is compared to that of the experimental carotid-femoral grafts. There was little similarity between the healing observed in the descending thoracic aortic and subcutaneous positions; the descending thoracic aorta grafts healed rapidly and completely by eight weeks, with thrombus-free flow surfaces covered with confluent endothelial cells. However, the inner wall of the carotidfemoral grafts remained largely unhealed at one year, with endothelialization largely confined to a limited pannus ingrowth across the anastomoses. This indicates that the implant site in the dog is a primary determinant of the rapidity and completeness of healing. The clinical axillofemoral grafts explanted after 20 months to six and one-half years exhibited healing characteristics that were similar to those seen in the carotid-femoral position in the dog. These findings suggest that valid comparisons of interspecies healing require data obtained from similar implant locations. They present a challenge to the previously held concepts of major interspecies differences in speed and extent of tissue ingrowth based on data obtained from noncomparable implant sites.
Similar content being viewed by others
References
BERGER K, SAUVAGE LR, RAO AM, WOOD SJ. Healing of arterial prostheses in man: its incompleteness.Ann Surg 1972;175:118–127.
SAUVAGE LR, BERGER KE, WOOD SJ, YATES SG II, SMITH JC, MANSFIELD PB. Interspecies healing of porous arterial prostheses; observation, 1960 to 1974.Arch Surg 1974;109:698–705.
WESOLOWSKI SA, FRIES CC, HENNIGAR G, FOX LM, SAWYER PN, SAUVAGE LR. Factors contributing to long-term failures in human vascular prosthetic grafts.J Cardiovasc Surg 1964;5:3–17.
DIDISHEIM P. Species selection for vascular graft evaluation. In: KAMBIC HE, KANTROWITZ A, SUNG P (eds). Vascular Graft Update: Safety and Performance. Philadelphia: American Society for Testing and Materials, 1986, p 174.
YATES SG, BARROS D'SA AAB, BERGER K, FERNANDEZ LG, WOOD SJ, RITTENHOUSE EA, DAVIS CC, MANSFIELD PB, SAUVAGE LR. The preclotting of porous arterial prostheses.Ann Surg 1978;188:611–622.
SAUVAGE LR, BERGER K, BEILIN LB, SMITH JC, WOOD SJ, MANSFIELD PB. Presence of endothelium in an axillary-femoral graft of knitted Dacron with an external velour surface.Ann Surg 1975;182:749–753.
MATHISEN SR, WU H-D, SAUVAGE LR, USUI Y, WALKER MW. An experimental study of eight current arterial prostheses.J Vasc Surg 1986;4:33–41.
SAUVAGE LR, BERGER K, WOOD SJ, NAKAGAWA Y, MANSFIELD PB. An external velour surface for porous arterial prostheses.Surgery 1971;70:940–953.
Author information
Authors and Affiliations
About this article
Cite this article
Durante, K.R., Wu, HD., Sauvage, L.R. et al. Implant site: A determinant of completeness of arterial prosthesis healing in the dog and possibly in humans. Annals of Vascular Surgery 4, 171–178 (1990). https://doi.org/10.1007/BF02001374
Issue Date:
DOI: https://doi.org/10.1007/BF02001374