Abstract
Background
Intermittent intraoperative neuromonitoring cannot prevent preparative surgical damage or predict imminent recurrent laryngeal nerve (RLN) damage with subsequent development of loss of electromyogram (EMG) signal during thyroid surgery. In case the nerve is stressed, i.e., from traction near the ligament of Berry, the nerve injury is only detected after it has occurred, not allowing the surgeon to correct the mechanical maneuver and salvage nerve function intraoperatively.
Methods
The unusual clinical scenario of sacrifice of a tumor-infiltrated RLN was used to study real-time evolution of RLN injury caused by mechanical distention. The ipsilateral vagus nerve (VN) was continuously stimulated with a new stimulation probe, and changes in EMG response were correlated with the varying levels of stretch and traction.
Results
Mechanical traction induced an intermittent depression of EMG amplitudes as a sign of impaired propagation of axon potentials or synaptic transmission. Prolonged mechanical stress caused a long-lasting depression of EMG response. When the mechanical distention was relieved, neurotransmission was gradually restored, with reappearance of singular muscle depolarization of full magnitude interspersed between the barely detectable deflections in the EMG recording. These responses of full amplitude appeared with increasing frequency, until the regular continuous EMG pattern was completely restored.
Conclusions
Only continuous VN stimulation serves to detect early changes in EMG response that indicate imminent danger to RLN functional integrity and alerts the surgeon to immediately correctable surgical actions, thus possibly preventing nerve damage or transforming damage into a reversible event.
Access this article
We’re sorry, something doesn't seem to be working properly.
Please try refreshing the page. If that doesn't work, please contact support so we can address the problem.
References
Hermann M, Hellebart C, Freissmuth M (2004) Neuromonitoring in thyroid surgery: prospective evaluation of intraoperative electrophysiological responses for the prediction of recurrent laryngeal nerve injury. Ann Surg 240:9–17
Randolph GW, Dralle H, Abdullah H et al (2011) Electrophysiologic recurrent laryngeal nerve monitoring during thyroid and parathyroid surgery: international standards guideline statement. Laryngoscope 121:1–16
Ulmer C, Koch KP, Seimer A et al (2008) Real-time monitoring of the recurrent laryngeal nerve: an observational clinical trial. Surgery 143:359–365
Schneider R, Przybyl J, Hermann M et al (2009) A new anchor electrode design for continuous neuromonitoring of the recurrent laryngeal nerve by vagal nerve stimulations. Langenbecks Arch Surg 394:903–910
Schneider R, Przybyl J, Pliquett U et al (2010) A new vagal anchor electrode for real-time monitoring of the recurrent laryngeal nerve. Am J Surg 199:507–514
Lorenz K, Sekulla C, Schelle J et al (2010) What are normal quantitative parameters of intraoperative neuromonitoring (IONM) in thyroid surgery? Langenbecks Arch Surg 395:901–909
Dralle H, Sekulla C, Haerting J et al (2004) Risk factors of paralysis and functional outcome after recurrent laryngeal nerve monitoring in thyroid surgery. Surgery 136:1310–1322
Chiang FY, Lu IC, Kuo WR et al (2008) The mechanism of recurrent laryngeal nerve injury during thyroid surgery—the application of intraoperative neuromonitoring. Surgery 143:743–749
Staal JA, Dickson TC, Gasperini R et al (2010) Initial calcium release from intracellular stores followed by calcium dysregulation is linked to secondary axotomy following transient axonal stretch injury. J. Neurochem 112:1147–1155
Nave K (2010) Myelination and support of axonal integrity by glia. Nature 468:244–252
Acknowledgments
The authors are grateful to Andreas Langer and Daniel Haberstroh from Dr. Langer Medical GmbH, Waldkirch, Germany, for technical and material support.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Schneider, R., Bures, C., Lorenz, K. et al. Evolution of Nerve Injury with Unexpected EMG Signal Recovery in Thyroid Surgery Using Continuous Intraoperative Neuromonitoring. World J Surg 37, 364–368 (2013). https://doi.org/10.1007/s00268-012-1853-0
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00268-012-1853-0