There is still a discussion whether cementless hip cup revisions should be performed with a press fit cup or a threaded ring implant.
Materials and methods
The results of 58 hip cup changes using the spherical, threaded ring “Munich” are presented. In 16 cases, the ring “Munich I” with a smooth surface and in 42 cases the ring “Munich II” with a corundum-blasted surface were implanted. Mean follow-up was 7.5 years; radiological signs of loosening and the Harris Hip Score were examined.
Results
The re-revision rate for aseptic loosening of the ring was 12/58 overall (Munich I: 5/16, Munich II: 7/42). The average Harris Hip Score was 76.2 overall (Munich I: 74.4, Munich II: 76.8). Higher survival rates were achieved with the “Munich II” ring due to improved construction details.
Conclusion
These rates are still below those reported for press-fit cups, the technique described by Slooff or reinforcement rings.
Similar content being viewed by others
Notes
Aesculap®
References
Aigner C (1998) 10 years results with the corundum-blasted Zweymuller titanium alloy threaded acetabular cup. Z Orthop Ihre Grenzgeb 136:110
Blaschke H, Kiss H, Dohnalek Ch, Abdolvahab F, Effenberger H, Dorn U (1998) Langzeitergebnisse der Hofer-Imhof Pfanne. In: Dorn U (ed) 10 Jahre Hofer-Imhof Pfanne, p 31
Blömer W, Ungetüm M (1992) Überlegungen zum Pfannendesign. Sphärisch oder konisch? Gewinde selbstschneidend oder geschnitten? In: Hipp E, Gradinger R, Ascherl R, Kölling A (eds) Die zementlose Hüftprothese. Demeter Verlag, Gräfelfing, p 137
Bonnomet F, Clavert P, Gicquel P, Lefebvre Y, Kempf JF (2001) Reconstruction by graft and reinforcement device in severe aseptic acetabular loosening: 10 years survivorship analysis. Rev Chir Orthop Reparatrice Appar Mot 87:135
Chareancholvanich K, Tanchuling A, Seki T, Gustilo RB (1999) Cementless acetabular revision for aseptic failure of cemented hip arthroplasty. Clin Orthop 361:140–149
Delaunay C, Cazeau C, Kapandji AI (1998) Cementless primary total hip replacement. Four to eight year results with the Zweymuller-Alloclassic prosthesis. Int Orthop 22(1):1–5
DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 121:20–32
Elke R, Wagner A, Berli B, Morscher E (2001) Acetabulum revision. Classifications and treatment possibilities. Orthopade 30:266
Engh CA, Griffin WL, Marx CL (1991) Cementless acetabular components. J Bone Joint Surg Br 72:53
Harris WH (1972) Clinical results using the Mueller-Charnley total hip prosthesis. Clin Orthop 86:95
Koudela K, Malotin T (1992) Reconstruction of the acetabulum during replacement of the aseptically loosened polyethylene cup. Acta Chir Orthop Traumatol Cech 68:162
Krüger-Franke M, Birk M, Rosemeyer B (1995) Results of cementless acetabular revision surgery. Z Orthop Ihre Grenzgeb 133: 374
More RC, Amstutz HC, Kabo JM, Dorey FJ, Moreland JR (1992) Acetabular reconstruction with a threaded prosthesis for failed total hip arthroplasty. Clin Orthop 282:114–122
Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9:33
Perka C, Ludwig R (2001) Reconstruction of segmental defects during revision procedures of the acetabulum with the Burch-Schneider anti-protrusio cage. J Arthroplasty 16:568
Schreurs BW, Slooff TJ, Buma P, Gardeniers JW, Huiskes R (1998) Acetabular reconstruction with impacted morsellised cancellous bone graft and cement. A 10- to 15-year follow-up of 60 revision arthroplasties. J Bone Joint Surg 80:391
Schreurs BW, Slooff TJ, Gardeniers JW, Buma P (2001) Acetabular reconstruction with bone impaction grafting and a cemented cup: 20 years’ experience. Clin Orthop 393:202–215
Stromberg CN, Herberts P (1996) Cemented revision total hip arthroplasties in patients younger than 55 years old. A multicenter evaluation of second-generation cementing technique. J Arthroplasty 11:489
van der Linde M, Tonino A (2001) Acetabular revision with impacted grafting and a reinforcement ring: 42 patients followed for a mean of 10 years. Acta Orthop Scand 72:221
Whaley AL, Berry DJ, Harmsen WS (2001) Extra-large uncemented hemispherical acetabular components for revision total hip arthroplasty. J Bone Joint Surg Am 83-A:1352
Winter E, Piert M, Volkmann R et al. (2001) Allogeneic cancellous bone graft and a Burch-Schneider ring for acetabular reconstruction in revision hip arthroplasty. J Bone Joint Surg Am 83-A:862
Wirtz DC, Niethard FU (1997) Etiology, diagnosis and therapy of aseptic hip prosthesis loosening—a status assessment. Z Orthop Ihre Grenzgeb 135:270
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
von Schulze Pellengahr, C., Düll, T., Müller, P.E. et al. Long-term results of 58 hip cup revision arthroplasties using a threaded ring implant. Arch Orthop Trauma Surg 127, 71–74 (2007). https://doi.org/10.1007/s00402-006-0187-5
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-006-0187-5