Abstract
Background
Comparison of different phacoemulsification parameters remains a frequently debated and methodologically challenging area. The focus of this study was twofold: (1) To suggest an accurate and simple in vivo model for comparison of phaco parameters with the use of femtosecond laser technology. (2) To compare parameters of intelligent phaco (IP) torsional and combination ultrasound modes using the proposed model.
Methods
Forty-three eyes underwent femtosecond laser-assisted cataract surgery (FLACS) with the precise division of each nucleus into two equal hemispheres. This approach was applied to compare two variants of phacoemulsification settings. In each case, one hemisphere was removed with the torsional ultrasound (US) with IP, while another hemisphere was evacuated with combination US. Aspiration time, cumulative dissipated energy (CDE), and fluid consumption were measured in all cases.
Results
Our approach allows for direct comparison of different phacoemulsification parameters on each half of the same nucleus under identical surgical conditions. When comparison between torsional US with IP and combination US was made, without regard to nucleus density, no statistically significant difference in CDE, aspiration time, and fluid consumption was found. However, in soft cataracts, CDE was lower with the use of torsional US with IP, while aspiration time and fluid consumption demonstrated no proven differences. In hard cataracts, CDE and aspiration time were significantly lower with the use of the combination US mode.
Conclusions
The suggested approach of testing different types of phaco settings on equal halves of the same nucleus can be easily applied for quantitative comparisons under identical surgical conditions. The data obtained suggest that IP technology is preferable when compared to a combination US mode for soft lenses while the combined US is more suitable for hard nuclei.
Similar content being viewed by others
References
Fakhry MA, El Shazly MI (2011) Torsional ultrasound mode versus combined torsional and conventional ultrasound mode phacoemulsification for eyes with hard cataract. Clin Ophthalmol 5:973–978
Helvacioglu F, Yeter C, Sencan S, Tunc Z, Uyar OM (2014) Comparison of two different ultrasound methods of phacoemulsification. Am J Ophthalmol 158:221–226
Zeng M, Liu X, Liu Y, Xia Y, Luo L, Yuan Z, Zeng Y (2008) Torsional ultrasound modality for hard nucleus phacoemulsification cataract extraction. Br J Ophthalmol 92:1092–1096
Liu Y, Zeng M, Liu X, Luo L, Yuan Z, Xia Y, Zeng Y (2007) Torsional mode versus conventional ultrasound mode phacoemulsification: randomized comparative clinical study. J Cataract Refract Surg 33:287–292
Kim DH, Wee WR, Lee JH, Kim MK (2010) The comparison between torsional and conventional mode phacoemulsification in moderate and hard cataracts. Korean J Ophthalmol 24:336–40
Mamalis N (2015) Which phacoemulsification modalities are the best? Comparing apples to apples. J Cataract Refract Surg 41:255–256
Demill DL, Zaugg BE, Pettey JH, Jensen JD, Jardine GJ, Wong G, Olson RJ (2012) Objective comparison of 4 nonlongitudinal ultrasound modalities regarding efficiency and chatter. J Cataract Refract Surg 38:1065–1071
Kirk KR, Ronquillo CJ, Jensen JD, Zaugg B, Barlow WR, Stagg BC, Pettey JH, Olson RJ (2014) Optimum on-time duty cycle for micropulse technology. J Cataract Refract Surg 40:1545–1548
Oakey ZB, Jensen JD, Zaugg BE, Radmall BR, Pettey JH, Olson RJ (2013) Porcine lens nuclei as a model for comparison of 3 ultrasound modalities regarding efficiency and chatter. J Cataract Refract Surg 39:1248–1253
Kim J, Ko D, Kim JY, Kim M, Tchah H (2013) Phaco-efficiency test and re-aspiration analysis of repulsed particle in phacoemulsification. Graefes Arch Clin Exp Ophthalmol 251:157–1161
Davison JA (2008) Cumulative tip travel and implied followability of longitudinal and torsional phacoemulsification. J Cataract Refract Surg 34:986–990
Han YK, Miller KM (2009) Heat production: longitudinal versus torsional phacoemulsification. J Cataract Refract Surg 35:1799–1805
Jun B, Berdahl JP, Kim T (2010) Thermal study of longitudinal and torsional ultrasound phacoemulsification: tracking the temperature of the corneal surface, incision, and handpiece. J Cataract Refract Surg 36:832–837
Sippel KC, Pineda R Jr (2002) Phacoemulsification and thermal wound injury. Semin Ophthalmol 17:102–109
Chen X, Yinghong J, Yi L (2013) Comparison of clear corneal incision injuries between torsional and conventional phacoemulsification. Graefes Arch Clin Exp Ophthalmol 251:2147–2154
Zacharias J (2015) Thermal characterization of phacoemulsification probes operated in axial and torsional modes. J Cataract Refract Surg 41:208–216
Zemba M, Cucu B, Furedi G, Enache V, Papadatu C, Ghigea B (2011) Intelligent phaco – always necessary? Oftalmologia 55:68–73
Tognetto D, Cecchini P, Leon P, Di Nicola M, Ravalico G (2012) Stroke dynamics and frequency of 3 phacoemulsification machines. J Cataract Refract Surg 38:333–342
Cionni RJ, Crandall AS, Felsted D (2011) Length and frequency of intraoperative occlusive events with new torsional phacoemulsification software. J Cataract Refract Surg 37:1785–1790
Acknowledgements
We are grateful to Fedor V. Konstantinov and Natalia V. Bondarenko for supporting this work. Valuable comments and suggestions of two anonymous referees significantly improved the paper.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Funding
No funding was received for this research.
Conflict of interest
All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the S. Fyodorov Eye Microsurgery Federal State Institution and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Rights and permissions
About this article
Cite this article
Tomilova, E., Shukhaev, S. A novel method to compare phacoemulsification parameters in vivo: two halves of one nucleus. Graefes Arch Clin Exp Ophthalmol 254, 1579–1584 (2016). https://doi.org/10.1007/s00417-016-3376-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-016-3376-0