Skip to main content
Log in

Jarvik 2000 biventricular assist device conversion from old pin-shaped bearing pumps to new conical bearing pumps

  • Case Report
  • Published:
Journal of Artificial Organs Aims and scope Submit manuscript

Abstract

A patient with a Jarvik 2000 (Jarvik Heart, Inc., New York, NY) biventricular assist device had massive hemolysis and device insufficiency of the right ventricular assist device (RVAD) after the anticoagulation was temporarily reversed due to cerebral hemorrhage. After the patient’s condition was stabilized by bridge use of CentriMag (Levitronix LLC, Waltham, MA) for right ventricle support, the Jarvik 2000 pumps with pin-shaped bearings were exchanged with new conical bearing pumps. Examination of the removed pumps revealed tight thrombus formation around the pin-shaped bearings.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Frazier OH, Myers TJ, Gregoric ID, Khan T, Delgado R, Croitoru M, Miller K, Jarvik R, Westaby S. Initial clinical experience with the Jarvik 2000 implantable axial-flow left ventricular assist system. Circulation. 2002;105:2855–60.

    Article  PubMed  CAS  Google Scholar 

  2. Kilic A, Nolan T, Li T, Yankey GK, Prastein DJ, Cheng G, Jarvik R, Wu ZJ, Griffith BP. Early in vivo experience with the pediatric Jarvik 2000 heart. ASAIO J. 2007;53:374–8.

    Article  PubMed  Google Scholar 

  3. Gibber M, Wu ZJ, Chang W, Bianchi G, Hu J, Garcia J, Jarvik R, Griffith BP. In vivo experience of the child-size pediatric Jarvik 2000 heart: update. ASAIO J. 2010;56:369–76.

    PubMed  Google Scholar 

  4. Saito S, Sakaguchi T, Sawa Y. Clinical report of long-term support with dual Jarvik 2000 biventricular assist device. J Heart Lung Transplant. 2011;30:845–7.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

This study was financially supported by the Health and Labour Sciences Research Grants from the Ministry of Health, Labour and Welfare of the Japanese government.

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Shunsuke Saito or Yoshiki Sawa.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Saito, S., Sakaguchi, T., Miyagawa, S. et al. Jarvik 2000 biventricular assist device conversion from old pin-shaped bearing pumps to new conical bearing pumps. J Artif Organs 16, 105–109 (2013). https://doi.org/10.1007/s10047-012-0669-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10047-012-0669-x

Keywords

Navigation