Abstract
Since the International Temperature Scale of 1990, ITS-90, was introduced, NPL has carried out a large number of calibrations of standard platinum resistance thermometers (SPRTs). Here we present an analysis of 876 calibrations of 435 SPRTs from 12 manufacturers in the various sub-ranges provided for in the ITS-90 between the triple point of argon (or, more usually, at the boiling point of nitrogen) to the freezing point of silver. Over 170 SPRTs have been calibrated more than once, in one case 10 times in 12 years. The calibration records have now been collated into a spreadsheet detailing source and identification data, calibration date, final value of \(R(0.01~{^{\circ }}\hbox {C})\), the measured values of resistance ratio \(W=R(t/{^{\circ }}\hbox {C})/R(0.01~{^{\circ }}\hbox {C})\) at the fixed points, and the coefficients of the deviation equations, a, b, c, and d. These data will permit investigations of the consistency over 23 years of the manufacture and quality of SPRTs including, inter alia, medium-to-long-term stability, ranges of W-values at each fixed point, correlations between W-values at different points and between the a and b coefficients, and the consistency between interpolations in different ITS-90 sub-ranges. A start has been made on these investigations, and this paper is a first report on the results.
Similar content being viewed by others
References
H. Preston-Thomas, The International Temperature Scale of 1990 (ITS-90), Metrologia 27 (1990) 3–10; erratum. Metrologia 27, 107 (1990)
I. Yang, C.H. Song, K.S. Gam, Y.-G. Kim, Long-term stability of standard platinum resistance thermometers in a range between \(0.01\;^{\circ }{\rm C}\) and \(419.527\;^{\circ }{\rm C}\). Metrologia 49, 803–808 (2012)
D.R. White, G.F. Strouse, Observations on sub-range inconsistency in the SPRT interpolations of ITS-90. Metrologia 46, 101–108 (2009)
D.R. White et al., Uncertainties in the realisation of the SPRT subranges of the ITS-90, BIPM Document CCT/08-19/rev (2009). http://www.bipm.org/cc/CCT/Allowed/24/D19_rev_WG3_Doc_rev_10July2009.pdf
R.L. Rusby, J.V. Pearce, C.J. Elliott, The link between SPRT sub-range inconsistency and Type 3 non-uniqueness in the ITS-90, in These Proceedings (2016)
BS EN ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories (Geneva, Switzerland: International Organization for Standardization)
K. Zhiru, L. Jingbo, L. Xiaoting, Study of the ITS-90 non-uniqueness for the standard platinum resistance thermometer in the sub-range \(0\;^{\circ }{\rm C}\) to \(419.527\;^{\circ }{\rm C}\). Metrologia 39, 127–133 (2002)
Acknowledgments
This work was funded by the UK National Measurement System for Engineering and Flow Metrology. J.C. gratefully acknowledges the Institute of Physics for supporting a Top 40 Work Placements Scheme summer placement at NPL in 2013. The authors wish to acknowledge all those who patiently carried out the calibrations over the last 25 years, notably Ann Duke, Jayne Gray, Stephen Bruce and Radka Veltcheva.
Author information
Authors and Affiliations
Corresponding author
Additional information
Selected Papers of the 13th International Symposium on Temperature, Humidity, Moisture and Thermal Measurements in Industry and Science.
Rights and permissions
About this article
Cite this article
Pearce, J.V., Crabb, J., Elliott, C.J. et al. Analysis of SPRT Calibration Data Obtained at NPL Since the Introduction of the ITS-90. Int J Thermophys 37, 124 (2016). https://doi.org/10.1007/s10765-016-2126-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10765-016-2126-1