Abstract
Background: Recently, the use of robotic assisted surgery has been utilized in cardiac surgical procedures. The use of robotics may offer benefits in precision, stability and control of instruments remotely. We report early experience with a novel remote robotic catheter control system (CCS).
Methods: We used a computerized robotically controlled catheter system that enables the user to remotely manipulate the tip of a catheter precisely in three dimensions. We tested the robotic catheter control systems ability to navigate within the heart and to make precise, rapid and repeatable movements. We compared the CCS with the ability of a standard quadripolar steerable ablation catheter placed in a deflectable sheath to navigate and make precision movements. Twelve ex-vivo porcine hearts were utilized to permit accurate measurements of navigation and precision. Eight targets were selected for navigation and precision testing. Time was measured for the catheter to reach the predefined target from a specific starting point to test navigation. In addition, time was measured to contact a discrete 0.8 mm target in order to test precision.
Results: The use of the CCS reduced the time needed for both navigation (8.5 ± 13.9 sec vs. 22.7 ± 26.7 sec, p = 0.002) and significantly decreased the time for precision targeting (10.1 ± 6.9 sec vs. 29.6 ± 26.4 sec, p < 0.001) in the specific RA and LA sites in the ex-vivo hearts.
Conclusions: The use of a computerized robotically assisted catheter control system is feasible and shows promise in rapid precision movement of the catheter. Further study is needed to elucidate the role of such a system in-vivo and in patient specific catheter ablation and mapping.
Similar content being viewed by others
References
Haissaguerre M, Shah DC, Jais P, Hocini M, Yamane T, Deisenhofer I I, Garrigue S, Clementy J. Mapping-guided ablation of pulmonary veins to cure atrial fibrillation. Am J Cardiol 2000;86(9 Suppl 1):K9–K19.
Haissaguerre M, Shoda M, Jais P, Nogami A, Shah DC, Kautzner J, Arentz T, Kalushe D, Lamaison D, Griffith M, Cruz F, de Paola A, Gaita F, Hocini M, Garrigue S, Macle L, Weerasooriya R, Clementy J. Mapping and ablation of idiopathic ventricular fibrillation. Circulation 2002;106(8):962–967.
Haissaguerre M, Shah DC, Jais P, Shoda M, Kautzner J, Arentz T, Kalushe D, Kadish A, Griffith M, Gaita F, Yamane T, Garrigue S, Hocini M, Clementy J. Role of Purkinje conducting system in triggering of idiopathic ventricular fibrillation. Lancet 2002;359(9307):677–678.
Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Metayer P, Clementy J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med 1998;339(10):659–666.
Faddis MN, Lindsay BD. Magnetic catheter manipulation. Coronary Artery Disease. 2003;14(1):25–27.
Faddis MN, Chen J, Osborn J, Talcott M, Cain ME, Lindsay BD. Magnetic guidance system for cardiac electrophysiology: A prospective trial of safety and efficacy in humans. J Am Coll Cardiol 2003;42(11):1952–1958.
Faddis MN, Blume W, Finney J, Hall A, Rauch J, Sell J, Bae KT, Talcott M, Lindsay B. Novel, magnetically guided catheter for endocardial mapping and radiofrequency catheter ablation. Circulation 2002;106(23):2980–2985.
Reger TB, Janhke ME. Robotic cardiac surgery. AORN Journal 2003;77(1):182–186.
Giulianotti PC, Coratti A, Angelini M, Sbrana F, Cecconi S, Balestracci T, Caravaglios G. Robotics in general surgery: Personal experience in a large community hospital. Archives of Surgery 2003;138(7):777–784.
Goh PM, Lomanto D, So JB. Robotic-assisted laparoscopic cholecystectomy. Surgical Endoscopy 2002;16(1):216– 217.
Gutt CN, Markus B, Kim ZG, Meininger D, Brinkmann L, Heller K. Early experiences of robotic surgery in children. Surgical Endoscopy 2002; 16(7):1083–1086.
Guyton SW, Robotic surgery: The computer-enhanced control of surgical instruments. Otolaryngologic Clinics of North America 2002;35(6):1303–1316, viii.
Hashizume M, et al. A new era of robotic surgery assisted by a computer-enhanced surgical system. Surgery 2002;131(1 Suppl):S330–5333.
Hollands, CM and Dixey LN, Applications of robotic surgery in pediatric patients. Surgical Laparoscopy, Endoscopy {&} Percutaneous Techniques 2002;12(1):71– 76.
Ballantyne GH. Robotic surgery, telerobotic surgery, telepresence, and telementoring. Review of early clinical results. Surgical Endoscopy 2002;16(10):1389–1402.
Davies B. A review of robotics in surgery. Proceedings of the Institution of Mechanical Engineers. Part H—Journal of Engineering in Medicine 2000;214(1):129–40.
Lanfranco AR, Castellanos AE, Desai JP, Meyers WC. Robotic surgery: A current perspective. Annals of Surgery 2004;239(1):14–21.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Al-Ahmad, A., Grossman, J.D. & Wang, P.J. Early Experience with a Computerized Robotically Controlled Catheter System. J Interv Card Electrophysiol 12, 199–202 (2005). https://doi.org/10.1007/s10840-005-0325-y
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s10840-005-0325-y