Abstract
A front-end for biopotential sensing in wearable medical devices has been recently proposed which is claimed to provide 100 GΩ input impedance by manually matching two resistor pairs in a positive- and a negative-feedback loop around an operational amplifier (op amp); the cost being that the equivalent input noise voltage doubles with respect to a simple non-inverting amplifier. The ECG acquired with capacitive (sic) electrodes through a cotton shirt is presented as a proof of the performance of the proposed circuit. It turns out, however, that the analysis ignores op amp’s input capacitance hence the effort to achieve a very high input resistance seems futile. Further, cotton is highly hygroscopic hence not an appropriate dielectric, so that there is no proof that the electrodes tested were actually capacitive. This comment addresses these two problems and some additional conceptual and methodological inaccuracies found in the paper.
Similar content being viewed by others
References
Wu R, Tang Y, Li Z, Zhang L, Yan F (2018) A novel high input impedance front-end for capacitive biopotential measurement. Med Biol Eng Comput 56:1343–1355. https://doi.org/10.1007/s11517-017-1781-0
Chi YM, Jung TP, Cauwenberghs G (2010) Dry-contact and noncontact biopotential electrodes: methodological review. IEEE Reviews Biomed Eng 3:106–119
Pallàs Areny R (2017) Comments on “capacitive biopotential measurement for electrophysiological signal acquisition: a review”. IEEE Sensors J 17:2607–2609
Fernandez M, Pallàs Areny R (2000) Ag-AgCl electrode noise in high-resolution ECG measurements. Biomed Instrum Tech 34:125–130
Huigen E, Peper A, Grimbergen CA (2003) Investigation into the noise of surface electrodes. Med Biol Eng Comput 40:332–338
Buxi D, Kim S, van Helleputte N, Altini M, Wijsman J, Yazicioglu RF, Penders J, van Hoof C (2012) Correlation between electrode-tissue impedance and motion artifact in biopotential recordings. IEEE Sensors J 12:3373–3383
Motchenbacher CD, Connelly JA (1993) Low-noise electronic design. John Wiley & Sons, New York
Chi YM, Maier M, Cauwenberghs G (2011) Ultra-high input impedance, low noise integrated amplifier for noncontact biopotential sensing. IEEE Journal on Selected Topics in Circuits and Systems 1:526–535
Author information
Authors and Affiliations
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Pallàs Areny, R. Comments on “A novel high input impedance front-end for capacitive biopotential measurement”. Med Biol Eng Comput 58, 267–269 (2020). https://doi.org/10.1007/s11517-019-02083-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11517-019-02083-x