In the past, clamping and crushing injuries occurred exclusively in therapy beds with scissor or joint lifting systems. This suggests an increased application risk for this type. The aim of the presented study is the systematic-analytical investigation of the clamping and crushing risk for choice types of electrically height-adjustable therapy beds. To analyse the application situation, 41 work process analyses and 100 structured interviews with employees were carried out in ten physiotherapy practices. Following, clamping and crushing risks were examined for four therapy bed-types with qualitative fault tree analyses (DIN EN 61025). Therapy beds with scissor or joint lifting systems show an increased risk of clamp and crush injuries. Therapy beds with lifting column systems have the lowest risk but are rarely used in practice due to their limited usability and high procurement costs. Prospective, technical safety measures are particularly necessary for therapy beds with scissors or joint lifting systems. New goods should be equipped with capacitive contact sensors in the lifting mechanism. For existing therapy beds, it is recommended to develop a locking box with improved usability (e.g. RFID technology), a three-step switch as well as the reversal of the direction of movement in case of all-round switches.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 39.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. Adegoke, B.O.A., Akodu, A.K., Oyeyemi, A.L.: Work-related musculoskeletal disorders among Nigerian physiotherapists. BMC Musculoskelet. Disord. 9(1), 112 (2008). https://doi.org/10.1186/1471-2474-9-112

    Article  Google Scholar 

  2. Girbig, M., Deckert, S., Druschke, D., et al.: Arbeitsbedingte Belastungen, Beschwerden und Erkrankungen von Physiotherapeuten in Deutschland: Ergebnisse einer Fokusgruppendiskussion. Physioscience 9(2), 66–71 (2013). https://doi.org/10.1055/s-0033-1335476

  3. Bundesinstitut für Arzneimittel und Medizinprodukte: BfArM-Bewertung bezüglich automatisch höhenverstellbarer Therapieliegen: Referenz-Nr.: 913/0704b (2004). https://www.bfarm.de/SharedDocs/Risikoinformationen/Medizinprodukte/DE/therapieliegen.html. Accessed 04 Sep 2019

  4. Stößlein, E.: Therapieliegen: Risikobewertung des BfArM, St. Augustin (2019)

    Google Scholar 

  5. Gerhards, M., Mehringer, B.: Mechanische Gefährdungen an elektrisch verstellbaren Therapieliegen: Eine sicherheitstechnische Betrachtung (2019). https://www.bgw-online.de/DE/Arbeitssicherheit-Gesundheitsschutz/Gefaehrdungsbeurteilung/Therapieliegen/Therapieliegen-Fachartikel.html. Accessed 04 Nov 2019

  6. Medizina: Sperrbox für die Therapieliege zum Nachrüsten (2019). https://www.medizina.de/sperrbox-therapieliege-nachruesten.html. Accessed 18 Nov 2019

  7. Berufsgenossenschaft für Gesundheitsdienst und Wohlfahrtspflege: Unfall bei Reinigungsarbeiten. BGW mitteilungen (2) (2017)

    Google Scholar 

  8. Perleth, M., Busse, R., Gerhardus, A., et al. (eds.): Health Technology Assessment: Konzepte, Methoden, Praxis für Wissenschaft und Entscheidungsfindung. Berliner Schriftenreihe Gesundheitswissenschaften. MWV Med. Wiss. Verl.-Ges, Berlin (2008)

    Google Scholar 

  9. U.S. Food & Drug Administration: Medical Device Reporting (MDR): How to Report Medical Device Problems (2019). https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. Accessed 04 Sep 2019

  10. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Patterson Medical Companies inc Performa Bar Activated 2-section Bobath Table, powered (2015). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=5044606&pc=INQ. Accessed 13 Nov 2019

  11. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Armedica MFG. Corp. Armedica Hi-Lo Treatment Table (2015). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=4663603&pc=INQ. Accessed 13 Nov 2019

  12. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Armedica MFG. Corp. Armedica Manufacturing Corp Hi-Lo Treatment Table (2015). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=4925834&pc=INQ. Accessed 13 Nov 2019

  13. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Armedica MFG. Corp. AMBA 240 (2015). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=5130875&pc=INQ. Accessed 13 Nov 2019

  14. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Armedica AM-BA Series Treatment Table (2015). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=4645290&pc=INQ. Accessed 13 Nov 2019

  15. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Armedica MFG. Corp. Armedica Hi-Lo Treatment Table (2014). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=4132582&pc=INQ. Accessed 13 Nov 2019

  16. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Patterson Medical Holdings Midland Hi-Lo Electric Table (2011). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=2384846&pc=INQ. Accessed 13 Nov 2019

  17. U.S. Food & Drug Administration: MAUDE Adverse Event Report: DJO Global TRT 600 Spine Therapy Table Powered Traction Table (2011). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=2129740&pc=INQ. Accessed 13 Nov 2019

  18. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Midland MFG Co./Patterson Medical Electric Hi-Lo Treatment Table (2010). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=1634980&pc=INQ. Accessed 13 Nov 2019

  19. U.S. Food & Drug Administration: MAUDE Adverse Event Report: Dynatronics Dynatron T3 Table Three Sectional Hi-Lo Treatment Table (2010). https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfMAUDE/detail.cfm?mdrfoi__id=1617360&pc=INQ. Accessed 13 Nov 2019

  20. Founten, L.: Clinic protocol blamed for infant crushed to death (2013). https://www.abc.net.au/news/2013-05-14/infant-massage-table-death-finding/4688060. Accessed 13 Nov 2019

  21. Mayring, P.: Qualitative inhaltsanalyse. In: Mey, G., Mruck, K. (eds.) Handbuch Qualitative Forschung in der Psychologie, 1st edn, pp. 601–613. VS Verlag für Sozialwissenschaften (GWV), Wiesbaden (2010)

    Chapter  Google Scholar 

  22. Deutsches Institut für Normung e. V.: Fehlerbaumzustandsanalyse 29.020(DIN EN 61025) (2007)

    Google Scholar 

  23. Himmelbauer, H., Treytl, A.: Fehlerbaumanalyse: Fault Tree Analyses (FTA) (1996)

    Google Scholar 

  24. Edler, F., Soden, M., Hankammer, R.: Fehlerbaumanalyse in Theorie und Praxis: Grundlagen und Anwendung der Methode. Springer, Heidelberg (2015). https://doi.org/10.1007/978-3-662-48166-0

    Book  Google Scholar 

  25. Karrock, L.: Die Risikoanalyse: Analysieren, erfassen, bewerten, 1st edn. Books on Demand, Norderstedt (2010)

    Google Scholar 

  26. Bundesinstitut für Arzneimittel und Medizinprodukte: Aktualisierte Empfehlung des BfArM: Fall-Nr. 0785/03 (2019). https://www.bfarm.de/SharedDocs/Risikoinformationen/Medizinprodukte/DE/therapieliegen_update.html. Accessed 19 Nov 2019

  27. Deutsches Institut für Normung e. V.: Medizinprodukte - Anwendung des Risikomanagements auf Medizinprodukte 11.040.01(DIN EN ISO 14971) (2009)

    Google Scholar 

  28. Gesetz über die Durchführung von Maßnahmen des Arbeitsschutzes zur Verbesserung der Sicherheit und des Gesundheitsschutzes der Beschäftigten bei der Arbeit: ArbSchG (1996)

    Google Scholar 

Download references

Acknowledgments

The present study was funded by the Berufsgenossenschaft für Gesundheitsdienst und Wohlfahrtspflege (BGW) in Germany.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Elisabeth Ibenthal .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2020 Springer Nature Switzerland AG

About this paper

Check for updates. Verify currency and authenticity via CrossMark

Cite this paper

Ibenthal, E., Sobkowicz, S., Backhaus, C. (2020). Investigation of Clamping and Crushing Injuries with Electrically Height-Adjustable Therapy Beds. In: Duffy, V. (eds) Digital Human Modeling and Applications in Health, Safety, Ergonomics and Risk Management. Posture, Motion and Health. HCII 2020. Lecture Notes in Computer Science(), vol 12198. Springer, Cham. https://doi.org/10.1007/978-3-030-49904-4_10

Download citation

  • DOI: https://doi.org/10.1007/978-3-030-49904-4_10

  • Published:

  • Publisher Name: Springer, Cham

  • Print ISBN: 978-3-030-49903-7

  • Online ISBN: 978-3-030-49904-4

  • eBook Packages: Computer ScienceComputer Science (R0)

Publish with us

Policies and ethics