Long-Term Stability of Image Quality Measurements for Two Digital Mammography Systems

  • Conference paper
Breast Imaging (IWDM 2012)

Part of the book series: Lecture Notes in Computer Science ((LNIP,volume 7361))

Included in the following conference series:

Abstract

Contrast-detail measurements were made at approximately weekly intervals for three months, for two full-field mammography systems with different types of detector. The measured threshold contrast values were found to be reasonably stable but with some random variation. The coefficient of variance was 8-10% for detail sizes 0.1 and 1.0mm, and 3-5% for detail sizes 0.25 and 0.5mm. The output of both X-ray sets was also monitored, and found to vary within ±1% of the mean. The variation in threshold contrast is likely to be mainly due to variation of noise in the CDMAM images. Care should be taken when setting baselines and acceptable limits, so that measured changes in threshold contrast that are of the order of ±10% of the mean are not wrongly interpreted as significant changes in performance of a digital mammography system.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 39.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. European Commission (EC): European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th edn. Office for Official Publications of the European Communities, Luxembourg (2006)

    Google Scholar 

  2. Young, K.C., Alsager, A., Oduko, J.M., Bosmans, H., Verbrugge, B., Geertse, T., Van Engen, R.: Evaluation of software for reading images of the CDMAM test object to assess digital mammography systems. In: Proceedings of SPIE Medical Imaging, 69131C, 1–11 (2008)

    Google Scholar 

  3. Young K.C., Cook J.J.H., Oduko J.M., Bosmans H.: Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systems. In: Proceedings of SPIE Medical Imaging, vol. 614206, pp. 1–13 (2006)

    Google Scholar 

  4. Young, K.C., Cook, J.J.H., Oduko, J.M.: Automated and Human Determination of Threshold Contrast for Digital Mammography Systems. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds.) IWDM 2006. LNCS, vol. 4046, pp. 266–272. Springer, Heidelberg (2006)

    Chapter  Google Scholar 

  5. Yang, C.-Y.J., Van Metter, R.: The variability of software scoring of the CDMAM phantom associated with a limited number of images. In: Proceedings of SPIE Medical Imaging, 65100C (2007)

    Google Scholar 

  6. Dance, D.R., Young, K.C., van Engen, R.E.: Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA dosimetry protocols. Phys. Med. Biol. 54, 4361–4372 (2009)

    Article  Google Scholar 

  7. Mackenzie, A., Dance, D.R., Workman, A., Yip, M., Wells, K., Young, K.C.: Development and validation of a method for converting images to appear with noise and sharpness characteristics of a different detector and X-ray system. Med. Phys. 39(5), 2721–2734 (2012)

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2012 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Oduko, J.M., Young, K.C. (2012). Long-Term Stability of Image Quality Measurements for Two Digital Mammography Systems. In: Maidment, A.D.A., Bakic, P.R., Gavenonis, S. (eds) Breast Imaging. IWDM 2012. Lecture Notes in Computer Science, vol 7361. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-31271-7_28

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-31271-7_28

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-31270-0

  • Online ISBN: 978-3-642-31271-7

  • eBook Packages: Computer ScienceComputer Science (R0)

Publish with us

Policies and ethics