Comparison of a New Brain Tissue Oxygenation Probe with the Established Standard

  • Chapter
  • First Online:
Intracranial Pressure and Brain Monitoring XIV

Part of the book series: Acta Neurochirurgica Supplementum ((NEUROCHIRURGICA,volume 114))

Abstract

Introduction: Continuous bedside brain tissue oxygenation (pbrO2) monitoring using the Licox system is an established method for detecting secondary ischemia in comatose patients with acute brain injury. The purpose of the current study was to compare the newly introduced Raumedic pbrO2 probe with the established standard.

Methods: Eighteen patients with acute traumatic brain injury or aneurysmal subarachnoid hemorrhage had pbrO2 probes of both types implanted side by side in the same vascular territory at risk of ischemia. Data were analyzed by the Bland–Altman method as well as random effect regression models to correct for multiple measurements per individual.

Results: Both types of probes were able to display spontaneous fluctuations of pbrO2 as well as reactions to therapy. Mean measurement difference between the Licox and Raumedic probes was −2.3 mmHg, with corresponding 95% limits of agreement of −32.3 to 27.5 mmHg. Regarding an ischemia threshold of 15 mmHg, both probes were in agreement in 78% and showed disparate results in 22%.

Conclusions: Our data suggest that the pbrO2 measurements of the two systems cannot be interchanged. Although we were unable to determine which system delivers more valid data, we do think that more rigorous testing is necessary before implementing the new probe in clinical routine.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 229.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 299.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 299.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. Ramakrishna R, Stiefel M, Udoetuk J, Udoteuk J, Spiotta A, Levine JM et al (2008) Brain oxygen tension and outcome in patients with aneurysmal subarachnoid hemorrhage. J Neurosurg 109(6):1075–1082

    Article  PubMed  Google Scholar 

  2. Narotam PK, Morrison JF, Nathoo N (2009) Brain tissue oxygen monitoring in traumatic brain injury and major trauma: outcome analysis of a brain tissue oxygen-directed therapy. J Neurosurg 111(4):672–682

    Article  PubMed  Google Scholar 

  3. Spiotta AM, Stiefel MF, Gracias VH, Garuffe AM, Kofke WA, Maloney-Wilensky E et al (2010) Brain tissue oxygen-directed management and outcome in patients with severe traumatic brain injury. J Neurosurg 113(3):571–580

    Article  PubMed  Google Scholar 

  4. Citerio G, Piper I, Cormio M, Galli D, Cazzaniga S, Enblad P et al (2004) Bench test assessment of the new Raumedic Neurovent-P ICP sensor: a technical report by the BrainIT group. Acta Neurochir (Wien) 146(11):1221–1226

    Article  CAS  Google Scholar 

  5. Citerio G, Piper I, Chambers IR, Galli D, Enblad P, Kiening K et al (2008) Multicenter clinical assessment of the Raumedic Neurovent-P intracranial pressure sensor: a report by the BrainIT group. Neurosurgery 63(6):1152–1158; discussion 1158

    Article  PubMed  Google Scholar 

  6. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310

    Article  PubMed  CAS  Google Scholar 

  7. Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8(2):135–160

    Article  PubMed  CAS  Google Scholar 

  8. Myles PS, Cui J (2007) Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth 99(3):309–311

    Article  PubMed  CAS  Google Scholar 

  9. Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15(2):85–91

    Article  PubMed  CAS  Google Scholar 

  10. Purins K, Enblad P, Sandhagen B, Lewén A (2010) Brain tissue oxygen monitoring: a study of in vitro accuracy and stability of Neurovent-PTO and Licox sensors. Acta Neurochir (Wien) 152(4):681–688

    Article  Google Scholar 

  11. Orakcioglu B, Sakowitz OW, Neumann JO, Kentar MM, Unterberg A (2010) Evaluation of a novel brain tissue oxygenation probe in an experimental swine model. Neurosurgery 67(6):1716–1722; discussion 1722–1723

    Article  PubMed  Google Scholar 

  12. Huschak G, Hoell T, Hohaus C, Kern C, Minkus Y, Meisel H (2009) Clinical evaluation of a new multiparameter neuromonitoring device: measurement of brain tissue oxygen, brain temperature, and intracranial pressure. J Neurosurg Anesthesiol 21(2):155–160

    Article  PubMed  Google Scholar 

Download references

Conflict of interest statement

We declare that we have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Stefan Wolf .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2012 Springer-Verlag/Wien

About this chapter

Cite this chapter

Wolf, S., Horn, P., Frenzel, C., Schürer, L., Vajkoczy, P., Dengler, J. (2012). Comparison of a New Brain Tissue Oxygenation Probe with the Established Standard. In: Schuhmann, M., Czosnyka, M. (eds) Intracranial Pressure and Brain Monitoring XIV. Acta Neurochirurgica Supplementum, vol 114. Springer, Vienna. https://doi.org/10.1007/978-3-7091-0956-4_30

Download citation

  • DOI: https://doi.org/10.1007/978-3-7091-0956-4_30

  • Published:

  • Publisher Name: Springer, Vienna

  • Print ISBN: 978-3-7091-0955-7

  • Online ISBN: 978-3-7091-0956-4

  • eBook Packages: MedicineMedicine (R0)

Publish with us

Policies and ethics